§ 880.5445 - Intravascular administration set, automated air removal system.

Copy with citation
Copy as parenthetical citation

Identification. An intravascular administration set, automated air removal system, is a prescription device used to detect and automatically remove air from an intravascular administration set with minimal to no interruption in the flow of the intravascular fluid. The device may include an air identification mechanism, software, an air removal mechanism, tubing, apparatus to collect removed air, and safety control mechanisms to address hazardous situations.

Classification. Class II (special controls). The special controls for this device are:

Provide an argument demonstrating that all reasonably foreseeable hazards have been adequately addressed with respect to the persons for whose use the device is represented or intended and the conditions of use for the device, which includes the following:

Description of the device indications for use, design, and technology, use environments, and users in sufficient detail to determine that the device complies with all special controls.

Demonstrate that controls are implemented to address device system hazards and their causes.

Include a justification supporting the acceptability criteria for each hazard control.

A traceability analysis demonstrating that all credible hazards have at least one corresponding control and that all controls have been verified and validated in the final device design.

Appropriate software verification, validation, and hazard analysis must be performed.

The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible.

Performance data must demonstrate the sterility of fluid path contacting components and the shelf life of these components.

The device must be designed and tested for electrical safety and electromagnetic compatibility (EMC).

Nonclinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:

Device system and component reliability testing must be conducted.

Fluid ingress protection testing must be conducted.

Testing of safety controls must be performed to demonstrate adequate mitigation of hazardous situations, including sensor failure, flow control failure, improper device position, device malfunction, infusion delivery error, and release of air to the patient.

A human factors validation study must demonstrate that use hazards are adequately addressed.

The labeling must include the following:

The device's air identification and removal response time.

The device's minimum air volume identification sensitivity.

The minimum and maximum flow rates at which the device is capable of reliably detecting and removing air.

Quantification of any fluid loss during device air removal operations as a function of flow rate.

LegalFix

Copyright ©2024 LegalFix. All rights reserved. LegalFix is not a law firm, is not licensed to practice law, and does not provide legal advice, services, or representation. The information on this website is an overview of the legal plans you can purchase—or that may be provided by your employer as an employee benefit or by your credit union or other membership group as a membership benefit.

LegalFix provides its members with easy access to affordable legal services through a network of independent law firms. LegalFix, its corporate entity, and its officers, directors, employees, agents, and contractors do not provide legal advice, services, or representation—directly or indirectly.

The articles and information on the site are not legal advice and should not be relied upon—they are for information purposes only. You should become a LegalFix member to get legal services from one of our network law firms.

You should not disclose confidential or potentially incriminating information to LegalFix—you should only communicate such information to your network law firm.

The benefits and legal services described in the LegalFix legal plans are not always available in all states or with all plans. See the legal plan Benefit Overview and the more comprehensive legal plan contract during checkout for coverage details in your state.

Use of this website, the purchase of legal plans, and access to the LegalFix networks of law firms are subject to the LegalFix Terms of Service and Privacy Policy.

We have updated our Terms of Service, Privacy Policy, and Disclosures. By continuing to browse this site, you agree to our Terms of Service, Privacy Policy, and Disclosures.